The current U. S. Environmental Protection Agency-approved method for enterococci (Method 1600) in recreational water is a membrane filter (MF) method that takes 24 hours to obtain results. If the recreational water is not in compliance with the standard, the risk of exposure to enteric pathogens may occur before the water is identified as hazardous. Because flow cytometry combined with specific fluorescent antibodies has the potential to be used as a rapid detection method for microorganisms, this technology was evaluated as a rapid, same-day method to detect enterococci in bathing beach waters. The flow cytometer chosen for this study was a laser microbial detection system designed to detect labeled antibodies. A comparison of MF counts with flow cytometry counts of enterococci in phosphate buffer and sterile-filtered recreational water showed good agreement between the two methods. However, when flow cytometry was used, the counts were several orders of magnitude higher than the MF counts with no correlation to Enterococcus spike concentrations. The unspiked sample controls frequently had higher counts than the samples spiked with enterococci. Particles within the spiked water samples were probably counted as target cells by the flow cytometer because of autofluorescence or non-specific adsorption of antibody and carryover to subsequent samples. For these reasons, this technology may not be suitable for enterococci detection in recreational waters. Improvements in research and instrument design that will eliminate high background and carryover may make this a viable technology in the future.
Skip Nav Destination
Article navigation
Research Article|
June 01 2007
A critical evaluation of a flow cytometer used for detecting enterococci in recreational waters
Dawn N. King;
1National Exposure Research Laboratory, U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr. Cincinnati, OH 45268, USA
Tel.: 513-569-7093 Fax: 513-569-7170; E-mail: [email protected]
Search for other works by this author on:
Kristen P. Brenner;
Kristen P. Brenner
1National Exposure Research Laboratory, U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr. Cincinnati, OH 45268, USA
Search for other works by this author on:
Mark R. Rodgers
Mark R. Rodgers
2National Risk Management Research Laboratory, U.S. Environmental Protection Agency, 26 W. Martin Luther King Dr., Cincinnati, OH 45268, USA
Search for other works by this author on:
J Water Health (2007) 5 (2): 295–305.
Citation
Dawn N. King, Kristen P. Brenner, Mark R. Rodgers; A critical evaluation of a flow cytometer used for detecting enterococci in recreational waters. J Water Health 1 June 2007; 5 (2): 295–305. doi: https://doi.org/10.2166/wh.2007.012b
Download citation file: