Analytical calibration data and linearity

Compoundatr (min)bLR (μg L−1)SlopeInterceptcR2dFexpeEF
Nap 3.11 0.150–200 0.1827 0.3083 0.9991 4.21 238 
Acy 4.21 0.15–200 0.2944 0.1185 0.9928 2.83 214 
Ace 4.76 0.15–200 0.1471 0.2304 0.9981 1.97 205 
Flu 4.98 0.10–150 0.2634 0.2171 0.9952 3.05 371 
Phe 5.49 0.150–200 0.3182 0.0711 0.9943 2.11 169 
Ant 6.18 0.20–200 0.5462 −0.0835 0.9918 4.02 110 
Flt 6.81 0.50–250 0.1607 0.0091 0.9991 2.61 268 
Pyr 7.43 0.15–200 0.0831 0.1426 0.9985 2.08 137 
BaA 7.96 0.10–150 0.1295 −0.0127 0.9976 1.86 114 
Chr 9.35 0.150–200 0.1938 −0.0318 0.9953 2.47 186 
BbF 9.82 0.10–150 0.2967 0.0629 0.9994 3.15 162 
BkF 11.24 0.10–150 0.1356 0.0094 0.9927 2.26 171 
BaP 11.81 0.10–150 0.1182 0.0341 0.9980 2.02 155 
DBA 12.57 0.10–150 0.0774 −0.1033 0.9995 1.88 94 
BgP 13.74 0.10–150 0.1121 0.0614 0.9991 2.43 192 
Ind 14.48 0.10–150 0.1507 0.0826 0.9959 3.50 128 
Compoundatr (min)bLR (μg L−1)SlopeInterceptcR2dFexpeEF
Nap 3.11 0.150–200 0.1827 0.3083 0.9991 4.21 238 
Acy 4.21 0.15–200 0.2944 0.1185 0.9928 2.83 214 
Ace 4.76 0.15–200 0.1471 0.2304 0.9981 1.97 205 
Flu 4.98 0.10–150 0.2634 0.2171 0.9952 3.05 371 
Phe 5.49 0.150–200 0.3182 0.0711 0.9943 2.11 169 
Ant 6.18 0.20–200 0.5462 −0.0835 0.9918 4.02 110 
Flt 6.81 0.50–250 0.1607 0.0091 0.9991 2.61 268 
Pyr 7.43 0.15–200 0.0831 0.1426 0.9985 2.08 137 
BaA 7.96 0.10–150 0.1295 −0.0127 0.9976 1.86 114 
Chr 9.35 0.150–200 0.1938 −0.0318 0.9953 2.47 186 
BbF 9.82 0.10–150 0.2967 0.0629 0.9994 3.15 162 
BkF 11.24 0.10–150 0.1356 0.0094 0.9927 2.26 171 
BaP 11.81 0.10–150 0.1182 0.0341 0.9980 2.02 155 
DBA 12.57 0.10–150 0.0774 −0.1033 0.9995 1.88 94 
BgP 13.74 0.10–150 0.1121 0.0614 0.9991 2.43 192 
Ind 14.48 0.10–150 0.1507 0.0826 0.9959 3.50 128 

atr is retention time. bLR is linear range. cR2 is the determination coefficient. dFexp is the ratio of residual variance to squared pure error, critical F(0.05, 42, 30) = 1. 79 is the critical value of F with (I − 3) = 42 and (I − L) = 30 degrees of freedom at 95% confidence level, where I is the number of calibration samples (45) and L the number of concentration levels (15). eEF is enrichment factor.