A commercial colorimetric indophenol (IP) method is used for determining monochloramine (NH2Cl) concentrations for process control in chloraminated public water systems and chloramine-related research. The NH2Cl – IP method excludes some quality control procedures typically included in drinking water methods and is not approved by the United States Environmental Protection Agency (U.S. EPA) for compliance monitoring. Therefore, the authors developed and validated a more complete NH2Cl–IP method, building on the commercial technique, as a candidate for future approval. During method development, temperature impact on color development was investigated. Color development time increased as temperature decreased. Below 20 °C, times needed for full color development were greater than those reported in the commercial method, reaching nearly three times longer at 5 °C. This observed temperature dependence also applies to free ammonia and free chlorine indophenol methods. To avoid measurement errors of samples analyzed below 20 °C, use of reaction times determined in this study is recommended for these indophenol methods.

  • Temperature dependence of the colorimetric indophenol method was evaluated.

  • Reaction times to achieve full color development are up to three times longer than those used in the commercial method that is commonly used to assess drinking water.

  • Use of reaction times determined by the authors is recommended to avoid underestimation of monochloramine and free chlorine measurements and variable free ammonia measurements.

Graphical Abstract

Graphical Abstract
Graphical Abstract

In public water systems (PWSs) using chloramines (i.e., combined chlorine) as a disinfectant, monochloramine (NH2Cl) is the preferred inorganic chloramine species, as compared to dichloramine or trichloramine, because NH2Cl has a greater (i) taste and odor threshold and (ii) stability (AWWA 2013). Fortunately, NH2Cl is the dominant inorganic chloramine species formed at typical pH values (i.e., greater than 7) and chlorine–to–ammonia–nitrogen (Cl2:N) mass ratios (i.e., less than 5:1) used in PWSs (Jafvert & Valentine 1992; Vikesland et al. 2001).

Independent of their regulatory compliance monitoring, PWSs that use chloramines sometimes analyze samples to determine NH2Cl concentrations with a commercial NH2Cl indophenol (NH2Cl – IP) method (Hach Company 2019a). In addition, they sometimes use a free ammonia indophenol (FreeNH3—IP) method to provide FreeNH3 concentrations that uses the same chemistry as the NH2Cl – IP method. The measured NH2Cl and FreeNH3 concentrations may be used to assess (i) effectiveness of process control (e.g., chloramine formation), (ii) maintenance of adequate secondary (i.e., distribution system) disinfectant residual, and (iii) maintenance of conditions in the distribution system to minimize the potential for nitrification (AWWA 2013). Therefore, ensuring accurate NH2Cl and FreeNH3 concentration measurements are of practical importance to chloraminated PWSs.

The commercial NH2Cl – IP method is not approved by the United States (U.S.) Environmental Protection Agency (EPA) for compliance monitoring in U.S. PWSs (40 CFR § 141.74 (a)(2)). Thus, those PWSs that (i) need to demonstrate compliance and (ii) wish to use the NH2Cl – IP method for process control to optimize NH2Cl residuals and minimize nitrification risk in the distribution system must use multiple methods.

The total chlorine N,N–diethyl–p–phenylenediamine colorimetric (TC–DPD) (APHA AWWA WEF 2017) method is commonly used for compliance monitoring by U.S. PWSs that disinfect with chloramines (AWWA Disinfection Committee 2018), but the TC–DPD method may overestimate disinfection efficacy because of positive interferences from organic chloramines that have no to poor disinfection capability (Donnermair & Blatchley 2003), resulting in an overestimation of the effective disinfectant concentration (i.e., NH2Cl). Compared with the TC–DPD method, the NH2Cl – IP technique has been demonstrated to specifically quantitate NH2Cl and is not impacted by the presence of dichloramine, trichloramine, or organic chloramines under drinking water conditions (Lee et al. 2007). Therefore, the NH2Cl – IP technique is of interest for assessing the effective disinfectant residuals at PWSs employing chloramine treatment.

At present, Hach Company is the sole provider of powdered reagents for the commercially available NH2Cl – IP method. Their powder pillow-based method is conducted by (i) adding their powdered reagent (Monochlor F) to a known volume of water sample, (ii) shaking the sample for a specified amount of time to dissolve the reagent, (iii) allowing the sample to develop color for a specified amount of time based on sample temperature, and (iv) analyzing the color development using a portable handheld colorimeter (λ = 610 nm) or benchtop spectrophotometer (λ = 655 nm) to determine the NH2Cl concentration (Hach Company 2019a). The powdered reagent contains a cyanoferrate that catalyzes a reaction between NH2Cl present in the sample and a substituted phenol in the powder to form an intermediate monoimine compound (Figure 1). The intermediate monoimine compound couples with excess substituted phenol to form a green-colored indophenol in direct proportion to the concentration of NH2Cl present in the sample (Hach Company 2019a). This series of reactions (Figure 1) is a modified version of the Berthelot reaction series (Harp et al. 2001). To accurately determine the NH2Cl concentration, it is important that the sample completes color development prior to analysis.

Figure 1

Schematic of indophenol method reactions.

Figure 1

Schematic of indophenol method reactions.

Close modal

Towards the goal of ultimately approving a compliance monitoring method to specifically and accurately determine the effective disinfectant concentration in chloraminated PWSs, U.S. EPA's Office of Research and Development (ORD) and Office of Water (OW) undertook development and validation of an EPA NH2Cl – IP method for determining NH2Cl concentration, building upon the commercial method. Following multi-laboratory validation and publication, the new method will be a candidate for approval by U.S. EPA as an alternative compliance monitoring method under the Agency's established method-approval process.

During U.S. EPA's NH2Cl – IP method development, the correlation between sample temperature and rate of color development was investigated and found to deviate from currently published information for the commercial method (Hach Company 2019a, 2019b, 2019c). To provide detailed guidance to water treatment and distribution system operators and state primacy agencies, studies were conducted at a series of water temperatures between 5 and 30 °C.

Although not the focus of the current study, the results presented herein may also apply to the FreeNH3—IP (Hach Company 2019b) and free chlorine indophenol (FreeCl2—IP) (Hach Company 2019c) methods because these methods use the same indophenol reaction chemistry. The FreeCl2—IP method was approved by U.S. EPA for compliance monitoring of FreeCl2 after it was deemed equally effective relative to the free chlorine DPD compliance monitoring method. In research, temperature is often an experimental variable that is tightly controlled. Therefore, ensuring that adequate time is provided for full color development is important. Consequently, the temperature dependence of reaction times reported in this study has a substantial practical implication to the drinking water community and researchers conducting chloramine-related research. The authors have shared these findings with the developer of the commercial method and are coordinating with them to ensure the use of consistent reaction times. Currently, the commercial method developer has included verbiage in their methods cautioning analysts to re-read samples after 5–10 minutes when sampling in cold weather to ensure full color development. In the future, commercial method manufacturers may reference the reaction times developed in the current research or the future U.S. EPA NH2Cl–IP method.

Sample preparation

All solutions were prepared in chlorine demand-free glassware (Summers et al. 1996). Prior to use, the glassware was soaked for 24 hours in a 20 mg Cl2 L–1 sodium hypochlorite (Fisher Chemical, NJ, USA) solution and subsequently rinsed in triplicate with distilled water. A 10,000 mg Cl2 L–1 FreeCl2 stock solution was prepared from laboratory-grade sodium hypochlorite and ultra-pure water (UPW, 18.2 MΩ·cm, Milli–Q, Millipore Sigma, MA, USA) and then standardized by spectrophotometry using the known hypochlorite ion (OCl) molar absorptivity (350 cm–1M–1 at 292 nm) (Margerum et al. 1994). A 10,000 mg N L–1 FreeNH3—nitrogen stock solution was prepared from certified ACS-grade ammonium sulfate salt (Fisher Chemical, NJ, USA) in UPW and adjusted to pH 8.3 with sodium hydroxide (NaOH). A 1,000 mg Cl2 L–1 NH2Cl stock solution was prepared using the FreeCl2 and FreeNH3 stock solutions in UPW with a target Cl2:N mass ratio of 4:1 and target pH above 8.3 to ensure that NH2Cl was the predominant chloramine species. The NH2Cl stock solution was mixed on a magnetic stirrer for 15 minutes to ensure full formation of NH2Cl from the combination of FreeCl2 and FreeNH3 stock solution doses. The NH2Cl stock solution was then standardized by spectrophotometry using the known NH2Cl molar absorptivity (445 cm–1M–1 at 245 nm) (Valentine et al. 1986). NH2Cl stock solutions were prepared fresh every 4 hours to minimize NH2Cl degradation and were standardized by spectrophotometry immediately prior to any use.

NH2Cl sample solutions for use in the study were prepared from the NH2Cl stock solution at 2.0 mg Cl2 L–1 in UPW, buffered with 0.2 mM potassium phosphate monobasic (Fisher Chemical, NJ, USA) and adjusted to pH 9.0 with sodium hydroxide (Fisher Chemical, NJ, USA) to ensure that NH2Cl was the predominant chloramine species. Experiments were conducted in both high- and low-ionic strength matrices. The high ionic strength (HIS) matrix was prepared by adding 1.40 g of sodium bicarbonate, 1.48 g of sodium sulfate, and 1.54 g of sodium chloride per liter of UPW in addition to the previously described phosphate buffer. The low ionic strength (LIS) matrix was prepared with UPW and the phosphate buffer only.

Experimental approach

The color development times were evaluated in triplicate at 5, 8, 10, 13, 15, 20, 25, and 30 °C in both HIS and LIS water matrices. Aliquots of the 2.0 mg Cl2 L–1 NH2Cl solution were transferred headspace-free into 16 mL amber glass vials. Three individual sample vials were used for each temperature tested. The vials were kept in a circulating water bath at the desired test temperature prior to analysis. The digital temperature display on the water bath was verified with a NIST–traceable thermocouple thermometer (Digi–Sense Traceable Lollipop Thermometer, Cole–Parmer, IL, USA). Prior to each trial, a sample blank was analyzed using the same kinetic program settings and temperature to provide a zero reference for the spectrophotometer. For each temperature tested, a new vial was removed from the water bath, then 6 mL of sample was pipetted to waste, leaving 10 mL of sample in the vial as specified in the commercially available NH2Cl – IP method (Hach Company 2019a). Pre-packaged, powdered Monochlor F reagent was then added to the vial. The sample was then agitated on a vortexer (Vortex Genie 2, Fisher Scientific, NJ, USA) for 20 seconds to ensure consistent mixing for each sample. The sample was subsequently transferred to a 1 cm quartz cuvette and inserted into a Thermo Scientific Peltier Control and Cooling Unit (Model # PCCU1, Thermo Scientific, MA, USA) fitted in a Thermo Nicolet Evolution UV–Vis 300 spectrophotometer (Thermo Scientific, MA, USA). The rate of color development, as measured by light absorbance at a wavelength of 655 nm, was recorded at 30-second intervals using kinetic software provided with the spectrophotometer. The elapsed time between adding reagent to the 10 mL NH2Cl sample and beginning the time course analysis in the spectrophotometer was noted and added to the reaction time data (typically 1 minute). The samples were maintained at the target temperatures using the Peltier temperature control unit throughout the kinetic program analysis. The sample temperature was verified with a NIST–traceable thermocouple thermometer prior to and immediately after each set of kinetic temperature measurements to verify that the temperature readout on the Peltier unit was accurate. Depending on the ambient temperature and relative humidity, it was determined that accumulation of condensation on the outside of the cuvette would interfere with absorbance readings, especially at the colder temperatures tested, so the sample compartment on the spectrophotometer was continuously purged with compressed air to prevent condensation from forming on the cuvette during analysis. The spectrophotometer kinetic program was run until full color development, defined as a change in absorbance readings less than 1% relative difference for five consecutive timesteps (i.e., 2.5 minutes).

Reaction time determination

An example color development kinetic curve is shown in Figure 2 (15 °C experiment). The rate of color development follows a sigmoidal pattern, the slope of the linear portion of which is directly proportional to the sample temperature. Greater temperatures had faster color development and correspondingly steeper slopes. Figure 2 is representative of all temperatures tested in that, at a given temperature, little to no difference in the rate of absorbance development or final absorbance value was observed between experimental trials or between the HIS versus LIS matrices at a given temperature, indicating the reproducibility and limited impact of ionic strength on color development rate.

Figure 2

Color development at 15 °C for the monochloramine indophenol method.

Figure 2

Color development at 15 °C for the monochloramine indophenol method.

Close modal

To determine the time required for full color development, 95% of the final absorbance value (as defined in the experimental approach) measured at 655 nm was selected for each experimental trial. This assumes a 5% error in determining reaction completion, which is generally acceptable for this type of method (Vogel et al. 1989). For each triplicate temperature trial conducted in both HIS and LIS matrices, the time corresponding to 95% of the final absorbance (t95) was calculated (Figure 3).

Figure 3

Temperature impact on 95% color formation time in both high- and low-ionic strength water matrices.

Figure 3

Temperature impact on 95% color formation time in both high- and low-ionic strength water matrices.

Close modal
Assuming the reagent is in excess and color formation is proportional to reacted NH2Cl (i.e., Beer's Law), color formation may be represented as a pseudo-first-order reaction (Equation (1)).
(1)
In Equation (1), t is time (minutes); Ft is the fraction of color formation relative to the final absorbance at time t; Ct is the unreacted NH2Cl concentration (mg Cl2 L–1) remaining in the sample; and k is the rate constant for the reaction of NH2Cl with the method reagents (min–1). For any specific Ft (e.g., Ft = 0.95, which represents 95% color formation) and using Equation (1), the corresponding time (i.e., t95) is proportional to k (Equation (2)).
(2)
The degree to which k changes with temperature can be described by the Arrhenius equation (Equation (3)), which is an empirical relationship commonly used to represent temperature dependence of the reaction rate constants (i.e., k) (Stumm & Morgan 1996).
(3)
In Equation (3), T is temperature (Kelvin, K), Ea is the activation energy of the reaction (J·mol–1); R is the ideal gas constant (J·K–1·mol–1), A is a constant, and k and t95 are as previously described. Because of the relationship of k to t95 from Equation (2), Equation (3) also represents the theoretical impact of temperature on t95. Equation (3) can be rearranged in the form of a linear equation where Ea/R is the slope, (ln 3 – ln A) is the intercept, ln t95 is the ordinate, and 1/T is the abscissa (Equation (4)).
(4)

Equation (4) was regressed using the observed t95 versus T data and results are presented in Figure 4. The observed R2 of 0.99 supports the hypothesis that the reaction between NH2Cl and the IP method reagents can be modelled as a pseudo-first-order relationship with an Arrhenius temperature dependence.

Figure 4

Arrhenius relationship for 95% color formation time.

Figure 4

Arrhenius relationship for 95% color formation time.

Close modal

Table 1 summarizes the upper 95% confidence limit t95 values determined from linear regression and compares them to the recommended reaction times reported in the commercial indophenol methods. The color development times determined in the current study are greater than those recommended by the three commercial indophenol methods at temperatures less than 20 °C. They are almost three times as great for 5 °C. The practical implication for water treatment practice is that NH2Cl concentrations measured following the color development time recommendations associated with the commercial NH2Cl – IP methods underestimate the actual NH2Cl concentration (or FreeCl2 concentration if using the FreeCl2—IP methods) for samples with temperatures below 20 °C.

Table 1

Impact of temperature on color development time for indophenol methods

Sample temperature [°C]Sample temperature [°F]Current research determined color development time [min]Commercial method reported color development time [min]a
41 28 10 
45 22 
48 17 
10 50 15 
12 54 12 
14 57 10 
16 61 
18 64 
20 68 
23 73 2.5 
25 77 
30 86 
Sample temperature [°C]Sample temperature [°F]Current research determined color development time [min]Commercial method reported color development time [min]a
41 28 10 
45 22 
48 17 
10 50 15 
12 54 12 
14 57 10 
16 61 
18 64 
20 68 
23 73 2.5 
25 77 
30 86 

aTimes are the same for the commercial monochloramine, monochloramine and free ammonia, and free chlorine methods (Hach Company 2019a, 2019b, 2019c).

Practical implications on monochloramine and free chlorine measurement

Even though U.S. EPA's research indicates that reaction times presented in the commercial indophenol methods are too short at temperatures less than 20 °C, it should be noted that results obtained using the shorter reaction times would underestimate the actual NH2Cl or FreeCl2 concentration. Therefore, NH2Cl or FreeCl2 concentration measurements determined using reaction times specified in the existing commercial methods would be conservative with regard to measuring disinfectant residual and thus generally be more protective of public health with regard to maintaining a disinfectant residual.

Accordingly, underestimation of NH2Cl or FreeCl2 concentrations may lead to unintended consequences in PWS operation. For instance, operators may unnecessarily implement strategies to increase the NH2Cl or FreeCl2 dose to meet predefined residual targets. The unnecessary increase may result in additional disinfection byproduct formation or potential exceedance of the maximum residual disinfectant level for NH2Cl or FreeCl2 of 4.0 mg Cl2 L–1.

When conducting PWS field sampling at ambient temperatures less than 20 °C, two factors may lead to water sample temperatures increasing prior to and during measurement. First, sample analysis may occur in a field vehicle, and second, physical handling of samples will tend to increase sample temperature. If the sample temperature increases after the sample temperature is measured to determine the appropriate reaction time (Table 1), the reaction times reported in the subject study would be conservative (i.e., the time required for complete color development would be less than that specified based on the original sample temperature because the actual temperature at time of analysis is greater). In practice and given the small (10 mL) sample volume used in the method, it is likely that one, or both, of the stated scenarios would occur in cold conditions (e.g., less than 10 °C in the field). Therefore, to avoid potentially long reaction times, analysts should measure sample temperature just prior to sample analysis. Note that this discussion applies to the use of powdered reagents (i.e., Monochlor F) and not the portable parallel analyzer planar cuvette method that is also commercially available from the same vendor (Hach Company 2015, 2016). The commercially available portable parallel analyzer warms the sample, adjusts reaction times for sample temperature, and determines the reaction time automatically.

In contrast to PWS compliance monitoring and process control sampling, chloramine-related laboratory research often controls experimental conditions for temperature, and measurements at temperatures less than 20 °C may be more likely (i.e., in a laboratory environment, samples are less subject to the warming described in the previous paragraph). Therefore, the longer reaction times determined in this study are particularly important to help ensure accurate measurements during laboratory-based research studies.

Practical implications on free ammonia measurement

The current study also has implications for FreeNH3 measurement by the FreeNH3—IP method. But, as opposed to the NH2Cl – IP or FreeCl2—IP methods, where using an inadequate color development time will bias the NH2Cl or FreeCl2 measurement low, providing inadequate times in the FreeNH3—IP method may bias the FreeNH3 measurement either low or high.

The FreeNH3—IP method uses two aliquots from a single sample to determine the FreeNH3 concentration by difference, and for both aliquots it is important to provide the minimum time for full color development. The first sample aliquot (A1) is prepared as per the NH2Cl – IP method, where Monochlor F reagent is added and allowed to react for the recommended time. Then A1 is used to blank the spectrophotometer or colorimeter. The second sample aliquot (A2) is used to measure the FreeNH3 concentration and is treated in two steps. First, FreeNH3 Chlorinating Reagent (sodium hypochlorite and sodium hydroxide) is added to A2, reacting FreeNH3 present in A2 into NH2Cl. Second, and as with A1, Monochlor F reagent is added to A2 and allowed to react for the recommended time. Then A2 is measured on the instrument previously blanked with A1. The increase in color from A1 to A2 is proportional to the sample's FreeNH3 concentration.

If full color development does not occur in either A1 or A2, the measured FreeNH3 will be biased, and the bias may be either higher or lower, depending on which aliquot(s) did not reach full color development. To understand the potential practical implications, it is important to also note that after full color development is reached, the developed color is reported stable for at least 15 minutes (Hach Company 2019a, 2019b, 2019c). This 15-minute minimum color stability time was also confirmed through experiments conducted as part of U.S. EPA's NH2Cl method development and validation. Therefore, it would be logical to assume an analyst may allow A1 or A2 or both to react up to 15 minutes longer than the minimum recommended time as there is no perceived impact.

Based on the previous discussion and the current study, four practical scenarios (Table 2) can be envisioned when sample temperatures are less than 20 °C and the current commercial FreeNH3—IP method is used. All scenarios assume a theoretical sample is measured at 15 °C and contains NH2Cl and FreeNH3 concentrations of 2.0 mg Cl2 L–1 and 0.25 mg N L–1, respectively. As previously stated, there is no perceived detriment in letting A1 or A2 react longer than the minimum prescribed time. But, if full color development has not occurred when using the commercial method times for samples below 20 °C, then allowing a longer time than the minimum commercial method time will impact results as color development will continue in A1 and A2 until reaching full color development. The four scenarios in Table 2 represent combinations of using two color development times at 15 °C (Figure 2): (i) the commercial method time of 6 minutes, meaning that full color development was not reached (Ft = 0.68) and (ii) the time of 9 minutes, as determined in the current study, meaning full color development is reached (Ft = 0.95) based on acceptable measurement error (Vogel et al. 1989).

Table 2

Implications of inadequate color development times on free ammonia measurement with the commercial free ammonia indophenol method

ScenarioIs full color development reached in sample aliquot before use?
Method measured free ammonia concentration [mg N/L]Percent of sample free ammonia concentration
A1A2
No No 0.17 68 
Yes No 0.06 24 
No Yes 0.35 140 
Yes Yes 0.24 96 
ScenarioIs full color development reached in sample aliquot before use?
Method measured free ammonia concentration [mg N/L]Percent of sample free ammonia concentration
A1A2
No No 0.17 68 
Yes No 0.06 24 
No Yes 0.35 140 
Yes Yes 0.24 96 

Theoretical sample contains 2 mg Cl2/L monochloramine and 0.25 mg N/L free ammonia at 15 °C.

A1 = first sample aliquot used as an instrument blank.

A2 = second sample aliquot used to measure free ammonia concentration.

No = indicates using the commercial method minimum recommended color development time of 6 minutes.

Yes = indicates using the subject research minimum recommended color development time of 9 minutes.

In Scenarios 1, 2, and 3 (Table 2), A1 and/or A2 use the commercial method's minimum recommended time of 6 minutes. Because at least one aliquot does not reach full color development, resulting measured FreeNH3 concentrations are biased and range from 24 to 140% of the actual FreeNH3 concentration. By comparison, using the 9-minute time of the current study for both A1 and A2 (Scenario 4, Table 2) results in a simulated FreeNH3 measurement that is 96% of the actual FreeNH3 concentration and within acceptable measurement error (Vogel et al. 1989). Overall, the provided example illustrates the potential variability in measured FreeNH3 concentration when using reaction times that are too short (i.e., Scenarios 1, 2, and 3) for full color development. The practical implications of this example are important for an operator using the commercial FreeNH3—IP method (below 20 °C) when measuring FreeNH3 for process control (e.g., chloramine formation or nitrification prevention). The variability in FreeNH3 measurement could lead to difficulty in meeting target FreeNH3 concentrations. Therefore, ensuring the minimum color development time is met is necessary to minimize unnecessary uncertainty in FreeNH3 measurements.

The temperature study reported herein was part of a larger development and validation study to produce a U.S. EPA NH2Cl–IP method for measuring NH2Cl concentrations in drinking water samples. The required reaction times for stable color development in the analyzed samples were determined at a series of temperatures ranging from 5 °C to 30 °C. Color development time increased as sample temperature decreased and could be described by an Arrhenius relationship. The appropriate color development times determined in the subject study were greater (sometimes significantly so) than those described in the three commercially available indophenol methods (NH2Cl–IP, FreeNH3–IP, and FreeCl2–IP) for temperatures less than 20 °C. Overall, to avoid underestimation of NH2Cl and FreeCl2 measurements and potentially variable FreeNH3 measurements in conditions less than 20 °C, it is recommended that reactions times determined in the subject study be used rather than those specified in the commercial NH2Cl–IP, FreeNH3–IP, or FreeCl2–IP methods.

The authors acknowledge Glynda Smith, Will Adams, and Nick Dugan with the U.S. EPA (Cincinnati, OH, USA) for their assistance and advice in this work, and Scott Tucker (Hach Company, Loveland, CO) for fruitful discussions concerning the indophenol methods. This work has been subjected to the United States Environmental Protection Agency's (Agency's) review and has been approved for publication. The views expressed in this manuscript are those of the authors and do not necessarily represent the views or policies of the Agency. Any mention of trade names, products, or services does not imply an endorsement by the Agency. The Agency does not endorse any commercial products, services, or enterprises.

All relevant data are included in the paper or its Supplementary Information.

APHA, AWWA, WEF
2017
Standard Methods for the Examination of Water and Wastewater
, 23rd edn.
American Public Health Association, American Water Works Association, Water Environment Federation
,
Washington, DC
.
AWWA
2013
Manual of Water Supply Practices M56: Nitrification: Prevention and Control in Drinking Water
(
Routt
J. S. J.
ed.).
American Water Works Association
,
Denver, CO
.
AWWA Disinfection Committee
2018
2017 Water Utility Disinfection Survey Report
.
AWWA Technical & Educational Council, American Water Works Association
,
Denver, CO
.
Donnermair
M. M.
Blatchley
E. R.
2003
Disinfection efficacy of organic chloramines
.
Water Research
37
(
7
),
1557
1570
.
Hach Company
2015
Method 10269, Ammonia, Free. Indophenol Method, 0.05 to 0.50 mg/L as NH3-N, Chemkey Reagents
.
Loveland, Colorado
.
Hach Company
2016
Method 10270, Chloramine (Mono). Indophenol Method, 0.04 to 4.60 mg/L as Cl2, Chemkey Reagents
.
Loveland, Colorado
.
Hach Company
2019a
Method 10171, Chloramine (Mono). Indophenol Method, 0.04 to 4.50 mg/L Cl2 (LR), Powder Pillows
.
Loveland, Colorado
.
Hach Company
2019b
Method 10200, Chloramine (Mono) and Nitrogen, Free Ammonia. Indophenol Method, 0.04 to 4.50 mg Cl2/L or 0.01 to 0.50 mg/L NH3-N
.
Loveland, Colorado
.
Hach Company
2019c
Method 10241, Free Chlorine USEPA Indophenol Method, 0.04 to 4.50 mg Cl2/L
.
Loveland, Colorado
.
Harp
D. L.
Weise
P.
Franklin
S.
2001
Controlling Chlorination of Wastewater and Chloramination of Drinking Water. US Patent 6,315,950
.
Jafvert
C. T.
Valentine
R. L.
1992
Reaction scheme for the chlorination of ammoniacal water
.
Environmental Science & Technology
26
(
3
),
577
.
Lee
W.
Westerhoff
P.
Yang
X.
Shang
C.
2007
Comparison of colorimetric and membrane introduction mass spectrometry techniques for chloramine analysis
.
Water Research
41
,
3097
3102
.
doi:10.1016/j.watres.2007.04.032
.
Margerum
D. W.
Schurter
L. M.
Hobson
J. L.
Moore
E. E.
1994
Water chlorination chemistry: nonmetal redox kinetics of chloramine and nitrite ion
.
Environmental Science & Technology
28
(
2
),
331
337
.
Stumm
W.
Morgan
J. J.
1996
Aquatic Chemistry: Chemical Equilibria and Rates in Natural Waters
, 3rd edn.
Wiley
,
Hoboken, New Jersey
.
Summers
R. S.
Hooper
S. M.
Shukairy
H. M.
Solarik
G.
Owen
D.
1996
Assess DBP yield: uniform formation conditions
.
Journal of the American Water Works Association
88
(
6
),
80
93
.
Vikesland
J. P.
Ozekin
K.
Valentine
R. L.
2001
Monochloramine decay in model and distribution system waters
.
Water Research
35
(
7
),
1766
1776
.
Vogel
A. I.
Furniss
B. S.
Hannaford
A. J.
Smith
P. W. G.
Tatchell
A. R.
1989
Vogel's Textbook of Practical Organic Chemistry
, 5th edn.
Longman Scientific & Technical
,
Essex
.